HN HP INPUT POWER Voltage ModelClass 2; 12 to 30 Vdc/24 Vac, 15 mA max. mA ModelClass 2; Loop powered 12 to 30 Vdc only, 30 mA max. OUTPUT Voltage Model3-wire, observe polarity mA Model2-wire, not polarity sensitive (clipped & capped) HUMIDITY HS Element†Digitally profiled thin-film capacitive (32-bit mathematics) U.S. Patent 5,844,138 Accuracy @ 25°C**±1%, 2%, 3%, or 5% (specify)@10 to 80% RH; Multi-point calibration, NIST traceable Reset Rate***24 hours Stability±1%@20 °C (68 °F) annually, for two years Hysteresis1.5% typical LinearityIncluded in accuracy spec. Temperature Coefficient±0.1% RH/°C above or below 25 °C (typical) Scaling0 to 100% RH TEMPERATURE OPTION Optional Temperature Transmitter Output Digital, 4 to 20 mA (clipped & capped) or 0-5/ 0-10 V output; accuracy ±0.5 °C (±1 °F) typical HN and HP Series probe type humidity transmitters are easy to install and exceptionally accurate. Their long-term stability and trouble- free serviceability make them among the best in the industry. The electronics are embedded inside the probe, protecting them from condensation-related failures. The thin-film capacitive HS sensor elements are factory calibrated using NIST traceable calibration equipment, eliminating the need for field calibration. Field replacement of the sensor element is a snap with the patented removable sensor, lowering costs and reducing downtime. Available † † † The HS sensing element has a 1-year warranty. The element is not a part of the 5-year product warranty. †† The CE mark indicates RoHS2 compliance. Please refer to the CE Declaration of Conformity for additional details. * One side of transformer secondary is connected to signal common, so an Isolation transformer or dedicated power supply may be required. ** Specified accuracy with 24 Vdc supplied power with rising humidity. RTD/Thermistors are not compensated for internal heating of product. *** Reset Rate is the time required to recover to 50% RH after exposure to 90% RH for 24 hours. Shielded cabling is required for conformance to EMC standards. Technical information is available from factory upon request or is available on our website: www.veris.com. EMC Conformance - CE Option: Low Voltage Directive 2014/35/EU and EMC Directive 2014/30/EU. EMC Special Note: Connect this product to a DC distribution network or an AC/DC power adaptor with proper surge protection (EN 61000-6-1 specification requirements). Specifications OPERATING ENVIRONMENT Operating Humidity Range0 to 100% RH non-condensing Operating Temp Range-40 to 50 °C (-40 to 122 °F) WARRANTY Limited Warranty5 years † AGENCY APPROVALS Sensor element Thin-film capacitive sensor element recovers from 100% saturation Flexibile Pendant and insertion versions for application flexibility Corrosion resistant Electronics are encapsulated in stainless steel probe to resist corrosion Compatibility Polarity insensitive two-wire 4 to 20 mA or 3-wire 0-5/0-10 Vdc versions…flexible systems compatibity Interchangable Fully interchangeable element to 1%, 2%, 3%, or 5% accuracy…no calibration Calibration free Calibration-free interchangeable NIST traceable HS element APPLICATIONS • HVAC control for improved comfort and energy savings • Museums, schools, printing shops, and other locations requiring humidity control • Facilitating compliance with ASHRAE standards for environmental control and indoor air quality 800.354.8556 | +1 503.598.4564 | sales@veris.com | intl@veris.com TM Humidity Monitoring | Specialty Humidity Sensors HN & HP SERIES Pendant and Insertion HQ0001793.F 0117POWER SUPPLY 24 Vdc BLK RED BLUE 0 to 10 V HUMIDITY ORANGE 0 to 10 V TEMP GREEN 0 to 5 V HUMIDITY YELLOW 0 to 5 V TEMP – + POWER SUPPLY 12 to 24 Vdc/24 Vac BLK RED BLUE 0 to 10 V HUMIDITY GREEN 0 to 5 V HUMIDITY ORANGE THERMISTOR ORANGE THERMISTOR – + POWER SUPPLY 12 to 24 Vdc DIGITAL CONTROL Temperature 4 to 20 mA Return (optional) Humidity 4 to 20 mA Return BLUE HUMIDITY BLUE HUMIDITY White/Gray 0 to 1 V Test Leads for use with voltmeter ORANGE Thermistor ORANGE Thermistor + – – POWER SUPPLY 24 Vdc only DIGITAL CONTROL Temperature 4 to 20 mA Return (optional) Humidity 4 to 20 mA Return BLUE HUMIDITY BLUE HUMIDITY White/gray 0 to 1 V test leads for use with voltmeter ORANGE TEMP ORANGE TEMP + – – 10" (254 mm) 3.5" (89 mm) 0.5" Dia. (13 mm) Foam Gasket 2.5" x 2.5" (64 mm x 64 mm) 8.0" (203 mm) 36" (914 mm) 10.0" 254 mm 0.5" Dia. 13 mm 36" 914 mm US or EUTemp. Sensor Type AccuracyNIST S = Standard C = CE T = Temp X = No Temp (Stop here) B = 100R Platinum, RTD C = 1k Platinum, RTD D = 10k T2, Thermistor E = 2.2k, Thermistor F = 3k, Thermistor G = 10k CPC,Thermistor H = 10k T3, Thermistor J = 10k Dale, Thermistor K = 10k with 11k shunt, Thermistor M = 20k NTC, Thermistor N = 1800 ohm TAC, Thermistor Q = 1uA/˚C, Linitemp R = 10k US, Thermistor S = 10k 3A 221, Thermistor T = 100k, Thermistor U = 20k “D”, Thermistor W = 10k T2 high accuracy, Thermistor Y = 10k T3 high accuracy, Thermistor Output M = 4 to 20 mA V = 0-5V/0-10 Vdc N = NIST 1%, & 2% only X = None 2%, 3%, 5% only 1 = 1% 2 = 2% 3 = 3% 5 = 5% N = RH Insertion P = RH Pendant Enclosure H OPTION Temp Cert Sensor TypeRange = Transmitter Temp Cert Blank = None 1 = 1pt Cal 2 = 2pt Cal OPTION 1 = -40 to 122 °F (-40 to 50 °C) 2 = 32 to 122 °F (0 to 50 °C) Humidity Transmitter Combination Humidity RTD/Thermistor Combination Blank = None 1 = 1pt cal* 2 = 2pt cal* A Example: (No Temp) HP2XVSX Example: (With Temp) H Stop Here N2XVSTC2 HN SERIES Dimensional Drawing HN/HP (0-5V/0-10V VERSIONS) Wiring Diagram HN/HP WITH RTD/THERMISTOR (0-5V/0-10V VERSIONS) Wiring Diagram HN/HP WITH RTD/THERMISTOR (4-20 mA VERSIONS) Wiring Diagram HP SERIES Dimensional Drawing HN/HP (4-20 mA VERSIONS) Dimensional Drawing ORDERING INFORMATION * Not available with W and Y high-accuracy thermistors. Humidity Monitoring | HN & HP Series HQ0001793.F 0117Next >